Digital pH Sensor Ceraliquid Cps41d

Product Details
Customization: Available
Display: Digital
Portable: Non Portable
Gold Member Since 2012

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Plant Area
501~1000 square meters
Management System Certification
ISO 9000, ISO 14001, QC 080000, GMP
Find Similar Products

Basic Info.

Model NO.
CPS41D
Appliance
Food Analysis, Medical Research
Certification
ISO9001, CE
Trademark
endress+hauser
Origin
China
Production Capacity
10000

Product Description

Memosens glass electrode for the chemical and life sciences industries

Ceraliquid CPS41D is the digital high performer for harsh chemical applications, media with low conductivity or a considerable content of organic solvents. The sensor is designed for fast response guaranteeing you high process safety even in applications with fast-changing media. Thanks to Memosens digital technology, CPS41D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.
 

      Benefits

  • Resistant to poisoning due to constant refilling of KCl bridge electrolyte and separate reference lead

  • Applicable at very low conductivities (= 0.1 µS/cm) thanks to liquid KCl electrolyte

  • Suitable for cleaning in place (CIP) and sterilization in place (SIP)

  • Perfectly suited for quickly changing media: Combination of liquid KCl electrolyte and ceramic diaphragm enables fast response time

  • Maximum process safety through non-contact inductive signal transmission

  • Enables predictive maintenance due to storage of sensor and process-specific data

  • Reduced operating costs due to minimized process downtime and extended sensor lifetime
     

    Field of application

    Media with very low conductivities or a considerable amount of organic solvents or alcohols and fast changing media:

  • Chemical industry

  • Power plants, e.g. boiler feed water

  • With ATEX, FM and CSA approval for use in hazardous areas.

  • Life sciences, e.g. water for injection

    Features and specifications
     

  • Measuring principle

    Glass electrode

  • Application

    Chemical industry, organic chemicals, power stations, food industry, laboratory measurements

  • Characteristic

    Analog or digital pH electrodes for process engineering with ceramic diaphragm
    and KCl liquid electrolyte

  • Measurement range

    pH 0 to 14

  • Measuring principle

    Liquid-KCl with ceramic diaphragm

  • Design

    All shaft lengths with temperature sensor
    Digital electrode with Memosens technology

  • Material

    Glass and ceramic

  • Dimension

    Diameter: 12 mm (0.47 inch)
    Shaft length: 120, 225, 360 and 425 mm
    (4.72, 8.86, 14.17 and 16.73 inch)

  • Process temperature

    0 to 135 °C (32 to 275 °F)

  • Process pressure

    1 to 11 bar abs (15 to 160 psi)

  • Temperature sensor

    NTC30k

  • Ex certification

    ATEX, FM, CSA

  • Connection

    Contactless, inductive, digital connection head with Memosens technology

  • Ingres protection

    IP68

  • Additional certifications

    3A, EHEDG

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.