Customization: | Available |
---|---|
Display: | Digital |
Portable: | Non Portable |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Ceraliquid CPS41D is the digital high performer for harsh chemical applications, media with low conductivity or a considerable content of organic solvents. The sensor is designed for fast response guaranteeing you high process safety even in applications with fast-changing media. Thanks to Memosens digital technology, CPS41D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.
Resistant to poisoning due to constant refilling of KCl bridge electrolyte and separate reference lead
Applicable at very low conductivities (= 0.1 µS/cm) thanks to liquid KCl electrolyte
Suitable for cleaning in place (CIP) and sterilization in place (SIP)
Perfectly suited for quickly changing media: Combination of liquid KCl electrolyte and ceramic diaphragm enables fast response time
Maximum process safety through non-contact inductive signal transmission
Enables predictive maintenance due to storage of sensor and process-specific data
Reduced operating costs due to minimized process downtime and extended sensor lifetime
Media with very low conductivities or a considerable amount of organic solvents or alcohols and fast changing media:
Chemical industry
Power plants, e.g. boiler feed water
With ATEX, FM and CSA approval for use in hazardous areas.
Life sciences, e.g. water for injection
Features and specifications
Glass electrode
Chemical industry, organic chemicals, power stations, food industry, laboratory measurements
Analog or digital pH electrodes for process engineering with ceramic diaphragm
and KCl liquid electrolyte
pH 0 to 14
Liquid-KCl with ceramic diaphragm
All shaft lengths with temperature sensor
Digital electrode with Memosens technology
Glass and ceramic
Diameter: 12 mm (0.47 inch)
Shaft length: 120, 225, 360 and 425 mm
(4.72, 8.86, 14.17 and 16.73 inch)
0 to 135 °C (32 to 275 °F)
1 to 11 bar abs (15 to 160 psi)
NTC30k
ATEX, FM, CSA
Contactless, inductive, digital connection head with Memosens technology
IP68
3A, EHEDG